The localised density of H₂O₂ in the effluent of a cold atmospheric pressure plasma jet determined by continuous-wave cavity ring-down spectroscopy

The data set comprises full cavity ring-down spectra and absorption coefficients obtained from on/off-resonance measurements, in order to determine the spatial distribution of H2O2 in the cold atmospheric pressure plasma jet kINPen-sci. Therefore, the plasma jet was operated with 3 slm Ar and 3000 ppm water, and was equipped with a gas curtain of 5 slm O2. To determine the effective absorption length, the H2O2 absorption was measured in radial direction. These radial fits had a Gaussian-like shape. By performing Abel inversion on the Gaussian fits, spatial distributions of H2O2 in x and y and in x and z direction were obtained.

Field Value
Group
Authors
Release Date 2023-07-05
Identifier00ed1e3d-dcc9-4d6a-b6b2-0f57850fc27f
Permanent Identifier (DOI) 10.34711/inptdat.727
Permanent Identifier (URI) https://www.inptdat.de/node/727
Is supplementing 10.3389/fphy.2023.1221181
Plasma Source Name
Plasma Source Application
Plasma Source Specification
Plasma Source Properties

Atmospheric pressure
Frequency: ~1MHz
Power: 1-3 W

Plasma Medium Name
Plasma Medium Properties

Feed gas: 3 slm Ar with 3000 ppm humidity
Gas curtain: 5 slm O2

Plasma Diagnostics Name
Plasma Diagnostics Properties

Diagnostics method: spatially resolved continuous wafe cavity ring-down spectroscopy (cw-CRDS) at a laser wavelength λ=1230 cm-1 (8.1 µm).

Plasma Diagnostics Procedure

The spatial resolution in radial direction was obtained after performing Abel inversion. To determine the spatial distribution the on/off-resonance method was employed.